- BrandНефтехимавтоматика АО БСКБ
- Country of manufactureRussia
According to MI 2418-97 "Classification and application of technical means of tests of the oil and oil products" registered by Gosstandart of Russia 15.12.97 penetrometr it is recommended for definition of a penetration of bitumens and other bituminous materials.
AUTOMATIC FUNCTIONS OF THE DEVICE:
- determination of size of a penetration with an accuracy ±0,05мм;
- display of information on the course and results of test on liquid crystal
- 4th lower case display;
- stop of process of test upon termination of the analysis with the sound alarm system;
- preservation and viewing to 9 results of analyses in memory.
NOT AUTOMATIC FUNCTIONS OF THE DEVICE:
- raising of a table of a penetrometr in two modes (quickly and slowly) on pressing of keys;
SPECIFICATIONS:
1. Penetration range: from 0 to 630 units of a penetration.
2. Movement measuring instrument error: ±0,05 mm.
3. Penetration time: from 1 to 3599 pages.
4. Delay time before performance of a penetration: from 0 to 3599 pages.
5. Speed of movement of a table: from 0,02 to 5,0 mm/sec.
6. The main sizes of needles and admissions on them conform to requirements of GOST 1440-78
7. Lump of moving parts (a plunger, freight and a needle with the holder) with which immersion of a needle (100,0±0,15) g and (200±0,2) g is made, including:
1) mass of a needle (2,50±0,05) g;
2) plunger (47,5±0,05) g;
3) freights (50,0±0,05) and (150,0±0,05) g.
8. Food parameters:
1) tension (220+22/-33)B;
2) frequency (50±1) of Hz;
3) power consumption, no more: 50 V • And.
9. Overall dimensions of the device, no more: 245х280х505 mm.
10. Weight, no more: 12 kg.
11. Warranty period of operation of the device: 12 months.
COMPLETENESS:
Name |
Quantity |
Note |
Device Lintel PN-10 of B |
1 |
|
Operation manual |
1 |
|
Passport |
1 |
|
Spare parts |
||
Insert fusible VP2-B 4A 250B |
1 |
|
Replaceable parts |
||
Needle for bitumens |
10 |
|
Freight, m=50g |
1 |
|
Freight, m=150g |
1 |
|
Accessories |
||
Level (L <400 mm) |
1 |
|
Core testing (diameter 3,2) |
1 |
|
Core tarirovochny (L=63mm) |
1 |
|
Cup (h=35mm) |
5 |
|
Cup (h=60mm) |
5 |
|
Support |
1 |
|
Bath of a penetrometr |
1 |
|
Case for bituminous needles |
1 |
The gages recommended for certification
Name of means |
Limits of measurements |
Class accuracy |
Appointment at certification of the device |
Note |
Micrometer smooth MK-25 |
(0-25) mm |
±0,004мм |
Measurement of the sizes of penetration needles and cones |
|
Caliper L-250 SHTS-11-250-0,05- KT 1 |
From 0 to 160 mm From 0 to 250 mm |
Class of accuracy 1. Error ± 0,05mm |
The same |
|
Tool microscope IMTsL 150h75B |
To 150 mm |
±0,003 mm |
The same |
|
Scales electronic HR-300 |
(0,01-310) |
Accuracy class special 1. Error of ±1,2 mg |
Measurement of mass of mobile parts of a penetrometr |
|
Tarirovochny cores |
(40,0±0,05) mm (50,0±0,05) mm (63,0±0,05) mm |
Klass1 ±0,05 mm |
Check of range of a penetration and condition of mobile system |
By the additional order |
Stop watch SOSPR-2b-2-000 |
(1-60) sec. (1-60) mines |
Class of accuracy 2. Error ±1,8с in 60 min. |
Check of counting of time |
|
Control samples * |
Check of repeatability of indications |
* as control samples choose products which are used at operation of the device