Description
Technical characteristics:
The UV-6 series of dual-beam spectrophotometers of the improved design, consisting of 4 models: Autonomous models: UV6100 with a width of bandwidth of 1.8 nanometers of UV6300 with a width of bandwidth of 1.0 nanometers; PC models: UV-6100PC with a width of bandwidth of 1.8 nanometers and UF-6300RS with a width of bandwidth of 1.0 nanometers. Other characteristics of 4 models are almost identical, except for bandwidth. 2 detectors measure a sample at the same time for optimization of process of measurements and improvement of accuracy. The device provides excellent measurements in the range from 190 nanometer to 1100 nanometers. The spectrophotometer is suitable for pharmaceutical, biochemical and clinical laboratories, and also usual applications as the quantitative analysis, kinetics, scanning of a range of multicomponent solutions, and also DNA / the proteinaceous analysis. The Windows PC the software does these tools universal, providing the highest productivity for carrying out measurements.
The UV-6 series of dual-beam spectrophotometers of the improved design, consisting of 4 models: Autonomous models: UV6100 with a width of bandwidth of 1.8 nanometers of UV6300 with a width of bandwidth of 1.0 nanometers; PC models: UV-6100PC with a width of bandwidth of 1.8 nanometers and UF-6300RS with a width of bandwidth of 1.0 nanometers. Other characteristics of 4 models are almost identical, except for bandwidth. 2 detectors measure a sample at the same time for optimization of process of measurements and improvement of accuracy. The device provides excellent measurements in the range from 190 nanometer to 1100 nanometers. The spectrophotometer is suitable for pharmaceutical, biochemical and clinical laboratories, and also usual applications as the quantitative analysis, kinetics, scanning of a range of multicomponent solutions, and also DNA / the proteinaceous analysis. The Windows PC the software does these tools universal, providing the highest productivity for carrying out measurements.
Model | UV-6100 | UV-6100PC | UV-6300 | UV-6300PC |
Range of lengths of waves | 190-1100 nanometers | 190-1100 nanometers | ||
The minimum width of a strip distinguished by a spectrometer | 1.8 nanometer | 1.8 nanometer | 1.0 nanometer | 1.0 nanometer |
Optical system | Two beams, gradation 1200 lines/mm | |||
Accuracy of control of wavelength | ±0.3 Nanometer | |||
Repeatability at control of lengths of waves | 0.2 Nanometer | |||
Scanning speed | Max. 3000 Nanometers/min. | |||
Photometric accuracy | ±0.3%T | |||
Photometric reproducibility | 0.15%T | |||
Photometric range | - 0.3-3A, 0-200% of T.O.-9999 Conc | |||
Diffused light | 0.05%T | |||
Stability | ±0.001A/h (@ 500 Nanometers) | |||
Display | 5" display (320x240) | 3,8" display (320x240) | 5" display (320x240) | 3,8" display (320x240) |
Flattening of the basic line | ±0.001A (200-1000 Nanometers) | |||
Chamber for samples | It is suitable for 100 mm a ditch with the optional holder | |||
Lighting source | galogenovy or deyteriyevy lamp | |||
Exit | USB port or parallel port (printer) | |||
Energy consumption | 220B/50gts or 110B/60gts | |||
Sizes (ShhDhV) | 600x450x200 mm | |||
Weight | 22 kg | 22 kg |
Contact the seller
Dual-beam UV-6 spectrophotometer